Lovibond MD 200 COD analyzer


Lovibond MD 200 COD analyzer


Optics  LEDs, interference filters (IF) and photo sensor in transparent sample chamber. Depending

on the version, up to 3 different interference filters are used. Wavelength specifications of

interference filters:

430 nm Δλ = 5 nm

530 nm Δλ = 5 nm

560 nm Δλ = 5 nm

610 nm Δλ = 6 nm

Wavelength Accuracy   ± 1 nm

Photometric Range        -2500 - 2500 mAbs

Photometric Accuracy   3 % FS (T = 20 °C – 25 °C)

Display backlit LCD (on keypress)

Interfaces          Infrared

Operation          Foil keypad

Auto – OFF        Yes

Calibration         Factory calibration and user calibration. Reset to factory calibration possible.

Internal Storage              internal ring memory for 16 data sets

Power Supply   4 batteries (AA)

Battery Life Time             approx. 53 hours

Clock    Real Time Clock and Date

Portability          Portable

Environmental Conditions           temperature: 5 - 40 °C rel. humidity: 30 - 90 % (non condensing)

IP Rating             IP 67

Compliance       CE

Dimensions       110 x 55 x 190 mm

Weight 455 g