Vertical Laminar Flow Cabinet LCB-V1100
Vertical laminar flow cabinet provides sample protection as purified air travels across the working zone of the cabinet in a vertical, unidirectional stream and leaves the main work chamber across the entire open front of the cabinet.
Most vertical laminar flow cabinets also have perforations at the back wall of the work zone which are designed to eliminate air turbulence and the possibility of dead air corners in the work zone.
Vertical Laminar Flow Cabinet Features:
1. Ergonomic control panel location, centered and angled down for an easy reach & viewing.
2. Brushed stainless steel work surface is for easy cleaning.
3. The non-direct light source design can effectively relieve the operator's visual fatigue under the condition of ensuring the illumination.
4. Intuitive and easy-to-use microprocessor controller with LCD display, graphical and digital display of the operating status of each function, such as filter life, UV life etc.
Reliability:
1. Front sliding sash interlock with fan and UV lamp.
2. UV sterilization with UV lamp timing function.
3. Front sash and side windows are made of UV-resistant tempered glass.
4. The host and sockets have independent capacitors to prevent current overload.
5. Memory function in case of power-failure.
Convenience:
1. HEPA filter and UV life indicators facilitate a timely replacement plan .
2. Two side glass windows installed for easy experiment observation.
3. Easy-to-change pre-filter installed could extend the HEPA filter’s lifetime.
4. Equipped with temperature sensor to show the real temperature during operation.
5. Waterproof socket provide convenience for small instruments to be used in the operating area.
Vertical Laminar Flow Cabinet Specifications: Model LCB-V1100
External Size (WxDxH)(mm) 1050x710x1850
Internal Size (WxDxH)(mm) 950x615x625
Work Surface Height 750mm
Control System Display Microprocessor controller with LCD display, display the operating status of each function
Interlock Front sliding sash interlock with fan and UV lamp
Life Indicator HEPA filter and UV life indicators facilitate a timely replacement plan
Airflow System Airflow Velocity 0.2~0.5 m/s, speed adjustable
Air filtration System Pre-Filter 1pcs, Polyester fiber,washable
HEPA Filter 1 pcs, Efficiency: > 99.995% at particle size of 0.3μm
Cleanliness Grade ISO Class 5 (formerly Class 100)
Partial air circulation Perforations at the back wall of the work zone which are designed to eliminate air turbulence and the possibility of dead air corners in the work zone.
Cabinet Construction Main Body Electro-Galvanized steel with antimicrobial powder coating
Work Surface 304 Stainless steel
Front Window Manual sliding sash: 5 mm UV resistant tempered glass with counter-balance system and ergonomic handle
Side window 5 mm UV resistant tempered glass
Electrical System UV Lamp 20W×1
With timing function from 0 to 999 minutes, Interlock with front sliding sash.
Emission of 253.7 nanometers for best sterilization effect.
LED Lamp 6.5W×1
Consumption 155W
Not include socket loading power consumption
Power Supply AC220V±10%, 50/60HZ; 110V±10%, 60Hz
Illumination ≥900Lux
Sound Emission ≤65dB(A)
Standard Accessory 1. Base Stand; 2. LED Lamp; 3. UV Lamp; 4. Waterproof Socket×2
Optional Accessory Gas Tap
Gross Weight (kg) 178
Shipping Size (W×D×H)(mm) 1200x930x1330