Automatic Kjeldahl Apparatus Specifications:
Model KA-9830I
Analytical Range 0.1~200 mgN
Accuracy(RSD) ≤0.5%(30mgN, 0.1mol/L H+)
Recovery Rate ≥99.5%
Input Mode of Titrant Concentration Manual input
Sample Analysis Time 3-4min/(cooling water temp 18℃)
Blank Deduction of Sample Automatic deduction
User-made Recipe 99 groups
Re-add Alkali During Distillation Process 0-99mL/ per time
Fertilizer Analysis Model (nitro) Reaction time 0-2 hours
Cooling Water Consumption 1.2L-3L/min
Connectable Fittings cooling water/reagent level
Data Storage 1 million sets of data
Printer 57cm thermal automatic paper cutting printer
The Digestive Tube Discharges Waste Automatically Automatic discharge
Steam Pressure Monitoring and Alarm 40kpa
Distillation Sample Temperature Monitoring and Alarm 50℃
Cooling Water Flow Monitoring and Alarm 0.8-10L/min
Steam Flow Control 70%-100%
Automatic Cleaning of Alkali Pipeline Automatic cleaning
Safe Alkali Adding Mode Equipped (0-999 sec)
Steam Boiler over-Temperature Alarm System 120℃
Steam Boiler with Dry Burning-resistant Protection Equipped
Safety Gate Switch Equipped
Digestive Tube Vacancy Switch Equipped
Power Supply AC220V, 50Hz
Heating Power 2KW
Size(W×D×H)(mm) 500x460x710
Shipping Size (W×D×H)(mm) 1050x650x1000
G.W. (kg) 30/60
Supporting Products
Digestion Furnace GDF Series+Waste Gas Collection Hood WGCH Series+Acid-base Neutralizing Apparatus ANA-1B